Abstract

An amperometric biosensor has been developed for the quantitative determination of urea in aqueous solution. The principle is based on the use of pH-sensitive redox active dissolved hematein molecule. The enzyme, urease (Urs), was covalently immobilized on a conducting copolymer poly ( N-3-aminopropyl pyrrole-co-pyrrole) film, electrochemically prepared onto an indium-tin-oxide (ITO)-coated glass plate. The covalent linkage of enzyme and porous morphology of the polymer film lead to high enzyme loading and an increased lifetime stability of the enzyme electrode. Amperometric response was measured as a function of concentration of urea, at fixed bias voltage of 0.0 V vs. Ag/AgCl in a phosphate buffer (pH 7.0). The electrode gives a linear response range of 0.16–5.02 m M for urea in aqueous medium. The response time is 40 s reaching to a 95% steady-state current value, and 80% of the enzyme activity is retained for about 2 months.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.