Abstract

The design, construction and performance characteristics of an easily removable interface to couple a gas chromatograph (GC) to an inductively coupled plasma mass spectrometer (ICP-MS) for trace element speciation is described. The interface is based on the use of a heated metallic tube where the last part of the capillary GC column is inserted. This metallic tube is connected to a metallic T-piece where a flow of argon carrier gas is introduced perpendicularly and externally to the metallic tube, creating a high velocity flow of intermediate sheath gas, which prevents condensation of the analytes eluting from the column on the walls of the T-piece or on the connection tubing to the ICP-MS torch. A flexible non-heated PTFE tube is used to connect the exit of the T-piece to the ICP-MS torch. This new interface has been applied to the separation and detection of different organometallic compounds of Se, S, Pb, Hg and Sn.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.