Abstract
The design, construction and performance characteristics of an easily removable interface to couple a gas chromatograph (GC) to an inductively coupled plasma mass spectrometer (ICP-MS) for trace element speciation is described. The interface is based on the use of a heated metallic tube where the last part of the capillary GC column is inserted. This metallic tube is connected to a metallic T-piece where a flow of argon carrier gas is introduced perpendicularly and externally to the metallic tube, creating a high velocity flow of intermediate sheath gas, which prevents condensation of the analytes eluting from the column on the walls of the T-piece or on the connection tubing to the ICP-MS torch. A flexible non-heated PTFE tube is used to connect the exit of the T-piece to the ICP-MS torch. This new interface has been applied to the separation and detection of different organometallic compounds of Se, S, Pb, Hg and Sn.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.