Abstract

To determine absorbed dose to water it is essential that the radiation quality of the x-ray beam is known accurately. The unique method of defining the beam quality is to acquire a detailed knowledge of the photon spectrum at the point of measurement in water, but this is impractical. The aim of this work was to investigate a quality index that uniquely defines the ratio of mass-energy absorption coefficients of water to air at 2 cm deep in water . Three parameters, HVL, mean energy at 2 cm deep in water and the ratio of the doses at 2 and 5 cm deep in water were investigated as functions of . The quality index that best defines is the ratio of doses at 2 and 5 cm depth in water.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call