Abstract
An algorithm for correction of the geometrical distortion in digital subtraction angiography (DSA) images was developed. Originally invented for 3D X-ray angiography, the algorithm was implemented in a computer program designed to fulfil the specific needs of stereotaxic DSA. The algorithm is based on transformation of an image of a grid from a distorted image back into its original pattern. The same transformation is then applied pixel-by-pixel to the angiographic images, which are acquired in direct conjunction with the grid image, without moving the gantry. The algorithm was tested in phantom studies and in the clinical situation with seven patients in ten examinations. Comparisons were made between co-ordinate determinations made on conventional full-size cut film and those performed on uncorrected and corrected DSA images, using 30- and 23-cm fields of view. With our method of measurement we could not shown any remaining geometric distortion in the corrected DSA images. This distortion correction can, if properly applied, be used for high-precision stereotaxic DSA.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.