Abstract

Massive-flow air samplers are being deployed around the world to collect aerosol samples for analysis of radioactivity as a result of nuclear tests and nuclear accidents. An aerosol wind tunnel capable of an 1100 m3 min−1 flow rate was built at Lovelace Respiratory Research Institute (LRRI) to test the sampling efficiency of these samplers. This aerosol wind tunnel uses a stationary air blender to enhance mixing, and therefore it achieves the required uniform distribution of wind speed and aerosol concentration in the test section. The test section of the wind tunnel has a cross section that is 4.3 m × 3.7 m. The aerosol wind tunnel was tested for performance in terms of distribution of wind speed, turbulent intensity, SF6 tracer gas concentration, and aerosol concentration. Test criteria consistent with U.S. Environmental Protection Agency (EPA) and American National Standards Institute (ANSI) standards were adopted as the guidelines for the aerosol wind tunnel. Additional criteria for aerosol wind tunne...