Abstract
A high flow rate four-stage impacfor was developed for the determination of aerosol concentration as a function of both particle size and time. The unit is very useful for long-term sampling intervals (24 hr) and for sampling very dusty atmospheres. Gas-borne particulate matter is collected out on four rotating drums, each with a collection surface area of about 10 sq in.; this allows a large quantity of materials to be collected without danger of particle build-up and blow-off. A particle size-collection efficiency calibration for the unit is presented together with experimental data on wall losses, surface coatings, and other important operational considerations.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.