Abstract

This paper describes an adjustable flow restrictor for use in gas-driven implantable infusion pumps, which is based on the resistance of a flow through a porous ceramic material. The flow inside the walls of a ceramic tube can be adjusted between 270 nl/min and 1260 nl/min by changing the flow path length in the ceramic over a distance of 14 mm. The long-term stability of the flow restrictor has been analyzed. A drift of -8% from the nominal value was observed, which lies within the required tolerance of ±10% after 30 days. The average time needed to change the flow rate is 40 s. In addition, the maximum adjustment time was 110 s, which also lies within the specification.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.