Abstract

This article presents a motion artifact (MA) cancellation technique for accurate photoplethysmography (PPG)-based heart rate (HR) estimation. The MA is canceled using two PPG signals, measured with closely placed red and green LEDs. The proposed technique utilizes the characteristics of the two PPG signals: the high correlation in MA and their different AC/DC ratios. These characteristics allow the MA to be canceled by an adaptive filter while preserving the AC components. In addition, the use of the sign-sign least mean square (SS-LMS) algorithm for the adaptive filter minimizes the hardware resource requirements. To validate the technique, a prototype was implemented and experiments were conducted with six subjects performing three types of movements: walking, running, and squatting. The proposed MA cancellation method significantly reduced the mean absolute error (MAE) in HR estimation, from 9.83 bpm to 1.48 bpm on average, compared to the conventional bandpass filtered green PPG.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.