An acoustic impedance bridge is described that is capable of measuring the input impedance of the human ear in the frequency range of 250 to 1250 Hz, with an accuracy of 3 degrees and 5 percent in phase and magnitude, respectively. In addition to static impedance, the bridge measures and records both in-phase and quadrature components (re an arbitrary reference) of changes in the ear's impedance due to the acoustic reflex. Assumptions upon which the use of the bridge is based are stated, the calibration procedure is described, and an example of the use of the bridge is given.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.