Abstract

To assess the accuracy of the Microlife 3AS1-2 blood pressure device in pregnancy and pre-eclampsia in a low-resource setting. Prospective validation according to the British Hypertension Society protocol. A total of 45 pregnant women were recruited from Kimberley Hospital (South Africa), of whom 15 had pre-eclampsia. The Microlife 3AS1-2 device achieved an overall B/A grade in pregnancy (including pre-eclampsia), passing the British Hypertension Society protocol requirements and achieving the International Organization for Standardization standard with a mean difference and SD of -3.8 ± 7.3 and -1.5 ± 6.2 mmHg for systolic and diastolic pressures, respectively. The Microlife 3AS1-2 device can be recommended for use in pregnancy, including pre-eclampsia. In addition, it fulfils the requirements stipulated by the WHO for an automated blood pressure device suitable for use in a low-resource setting. This makes it the ideal device for antenatal clinics and primary healthcare facilities in low-income and middle-income countries.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.