Abstract

This paper describes the assessment of a method for measuring the oxygen consumption of infants and young children that is suitable for use in the cardiac catheterization laboratory. It is based upon the open-circuit technique and permits the continuous display of the oxygen consumption during the investigative procedures. The accuracy of the apparatus was assessed over the range 40-200 ml/min, both in vitro and in vivo. It was found that the apparatus was free from systematic error and that it had a random error of approximately 2% of the measured value.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.