Abstract

The design and performance of an accurate flow system utilizing critical velocity orifices is described. Results indicate that an absolute accuracy of three parts in a thousand is realized over a flow range of 0.002 to 2 cubic feet per minute. Calibration of the flow system was accomplished by means of a specially modified and carefully standardized wet test gas meter. Descriptions of the components of the flow system and of subsidiary calibrating equipment are presented. Critical flow orifice discharge coefficients are found to be independent of the type of gas; however, because of gas impurities, prediction of relative rates of flow of various gases through an orifice by use of theoretical relationships is uncertain, and experimental data are given to illustrate this.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.