Abstract

An accurate flowmeter based on the reduced constant-pressure method was devised for admitting the sample gas into the gauge calibration system. This flowmeter consists of a U-tube manometer, an accurate micrometer for measuring the gas pressure, and a diaphragm gauge for checking the deviation of the gas pressure. The gas pressure was kept constant by raising the mercury level in the U-tube manometer. The flow rate could be obtained from the gas pressure, the cross-sectional area of the U-tube manometer, and the rise-up rate of mercury level. With the U-tube manometer of 34. 0±0.1 mmφ in diameter, the flow rate of 10-6 to 10-2 Torr·l/s could be easily measured with an accuracy better than 1%. An actual example in which this flowmeter was used for the gauge calibration was presented.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.