Abstract

This study describes a relatively simple mechanical technique for measuring occlusal wear at the margin of dental restorations. The device was essentially a custom profilometer which consisted of a linear variable differential transformer with its core supported by double cantilever springs, a specially shaped stylus, a motor driven microscopic stage, custom computer software utilizing a colour coding to separate differences in slope, and a cross-hair-shaped mouse cursor to differentiate the boundary of wear. Two sets of visual standard stone casts were evaluated. The first set consisted of 17 stone replicas of machined steel dies and the second set consisted of five clinical casts, which have been used to measure the wear of composite restorations by direct vision. The reproducibility found when measuring the stone replicas of the machined steel dies was 2 microns, while, for the clinical standards, it was 16 microns.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.