Abstract

In this study, an eco-friendly and accurate analytical protocol was proposed for the determination of cadmium at trace levels. Prior to analysis in flame atomic absorption spectrophotometry (FAAS), an ultraviolet based photochemical vapor generation (PVG) method as a sample introduction system was employed to eliminate matrix effects. PVG system was combined with an online preconcentration technique called T-shaped slotted quartz tube-atom trapping (T-SQT-AT) for online preconcentration of cadmium in order to further increase analytical performance of the system. Hydrogen was used in the releasing of collected Cd atoms from the quartz surface. Under optimum conditions, limit of detection (LOD) and limit of quantitation (LOQ) of the developed method were specified as 2.5 ng/mL and 8.2 ng/mL, respectively. Percent recovery results for tap water and daphne tea samples were found to be between 87.6–105.3% and 90.9–103.3%, respectively.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.