Abstract

An ac quantum voltmeter based on a 10 V programmable Josephson array that is simple to use, provides dc and ac calibration up to kHz range for equipment widely used in metrology, and ensures direct traceability to a quantum-based standard, is developed. This ac quantum voltmeter is proven to match conventional Josephson standard systems at dc and extends its advantages up to 10 kHz in the low-frequency ac range. The ac quantum voltmeter is capable of performing calibrations up to 7 VRMS in the frequency range from dc to 10 kHz completely under software control. A direct comparison at dc has demonstrated an uncertainty better than 2 parts in 1010 (k = 2). The uncertainty at 1 kHz is better than 1.7 µV V−1 (k = 2) for a measurement time of 1 min. The ac quantum voltmeter is a robust and practical system that fulfils the needs of general metrology laboratories for quantum-based voltage calibrations.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.