Abstract
A simple and sensitive liquid chromatographic assay for simultaneous quantitation of amrinone and N-acetylamrinone in human plasma was developed. The method involves extraction of samples via activated solid-phase extraction Bond Elut C 18 disposable columns, followed by chromatographic separation on a reversed-phase phenyl column using isocratic condition and UV detection. The assay can measure concentrations of both compounds over the range 0.075-10 μlg ml −1. The injection interval is 11 min. The inter-day relative standard deviation (RSD) for replicate analysis of spiked samples is less than 10% and the accuracy more than 94% for both compounds over the standard curve range. The assay has been successfully applied to pharmacokinetic studies in humans.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.