Abstract
The flow of a saline-glycerin solution with sand particles through a continuous in vitro flow system was imaged by using two commercially available Doppler color flow mapping systems in a power mode (Toshiba SSH-160A and Advanced Technology Laboratories [ATL] Ultramark 9). The images generated from seven solutions with particle concentrations ranging from 0.0001 × 1012to 6 × 1012particles/liter and a mean velocity of 30 cm/s measured with use of pulsed Doppler ultrasound were used to examine the dependence of the power mode on particle concentration.To examine the velocity dependence, 20 mean velocities ranging from 0.1 to 0.53 m/s (3 to 30 liters/min) and three particle concentrations (1, 3 and 6 × 1012particles/liter) in the solution were used. The recorded images were digitized and analyzed off-line. The SUM values, or the adjusted color intensity levels in delineated areas of interest in the displayed flow, were compared. In general, the power mode was sensitive in displaying slower velocity flows; in the selected particle concentration and velocity ranges, it was both velocity and concentration dependent. The specific dependence differed for the two color flow mapping systems.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.