Abstract

Transcatheter closure of patent ductus arteriosus (PDA) by Amplatzer duct occluder is the treatment of choice. However, closure of very large ducts in infants with low weight is a challenge for the interventionalist because a large device may obstruct the aorta or left pulmonary artery. Difficulty is also encountered in advancing the device around the curve of the right ventricular outflow tract toward the pulmonary artery; this curve is tight, more or less at a right angle in infants, leading to kinking of the sheath, which increases fluoroscopic time. This is the first reported case of a very large PDA (8.7 mm), larger than the aorta (8.2 mm), successfully closed by an Amplatzer angled duct occluder in an infant weighing 5 kg.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.