Abstract

Amperometric gas sensors in a three-electrode configuration were investigated for hydrogen measurements in atmospheres containing carbon monoxide. The hydrogen-sensing working electrode was based on a platinum black electrode deposited on a porous Teflon ® substrate. To achieve selectivity, an FEP membrane was mounted at the gas inlet of the sensor. Hydrogen readily diffused through the FEP membrane and was detected on the Pt working electrode. The cross-sensitivity to gases such as carbon monoxide, ammonia and hydrogen sulfide was eliminated with nearly 100% efficiency in the short term. In addition, the poisoning effect of the carbon monoxide was virtually eliminated. Sensor stability to high levels of hydrogen was improved by the development of a hydrogen-insensitive reference electrode.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.