Abstract

A glassy carbon electrode (GCE) modified with Pd/IrO 2 provides excellent electrocatalytic oxidation of hydrogen peroxide. Glucose oxidase (GOD) and xanthine oxidase (XOD) were co-immobilized on the modified electrode with a thin film Nafion coated on the enzyme layer to form a glucose (Glu)/hypoxanthine (Hx) sensor, without interference from electroactive species such as ascorbic acid (AA) and uric acid (UA). Its response was evaluated with respect to the enzyme amount on the electrode, pH and temperature of the electrolyte. The prepared bienzymic biosensor, used as the detector of HPLC gave a detection limit of 1.0×10 −6 mol l −1 Glu and 2.0×10 −7 mol l −1 Hx (Hx) with a linear concentration range of 5.0×10 −6–2.5×10 −3 mol l −1 and 1.0×10 −6–5.0×10 −4 mol l −1, respectively. Coupled with microdialysis, it was used to monitor the concentrations of Glu and Hx in rat brain.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.