Abstract

A novel enzymeless glucose sensor of poly (sulfosalicylic acid) film modified nickel disk-microelectrode based on glucose electrochemical oxidation directly in 0.01 mol dm−3 NaOH has been fabricated. It shows excellent electrocatalytic activities towards glucose oxidation at a low positive potential of 0.49 V (vs. Ag/AgCl). The calibration curve is linear in the glucose concentration range of 6.0 × 10−6 to 1.0 × 10−3 mol dm−3 with a detection limit 9 × 10−7 mol dm−3 (S/N = 3) and sensitivity 3.8 mA μM−1 cm−2. The presented glucose sensor is selective, sensitive, and rapid in response time. It had been used to direct determination of glucose in serum.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.