Abstract
AbstractAn inexpensive, robust, and easily mechanized amperometric glucose senor based on a biocomposite material has been constructed. The biocomposite is made of graphite, a nonconductive epoxy resin, tetrathiafulvalene (TTF) as a redox mediator, and glucose oxidase (GOD). The enzyme retains its bioactivity on confinement in the rigid epoxy‐graphite matrix, with the oxidation current of the reduced mediator serving as the analytical signal (at 150 mV [vs. Ag/AgCl] in a pH 7.00 buffered solution featuring 0.1 M phosphate and 0.1 M KC1). The resulting biosensor shows a linear response in the 0.1 to 2 mM glucose range and produces a steady‐state signal within a few seconds (11 seconds for a 95% response). Simple polishing procedures can be used to regenerate a freshly bioactive transducer surface.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.