Abstract

A detection system for plasma and urinary oxalate involving a postcolumn enzymatic reaction and electrochemical detection is described. Oxalate oxidase was immobilized on AF-Tresyl Toyopearl 650 gel. The immobilized oxalate oxidase was packed into a stainless-steel column (10 x 4 mm I.D.) and used on-line as an immobilized enzyme reactor (IMER). The hydrogen peroxide produced by enzymatic reaction was detected amperometrically at a platinum electrode maintained at +0.5 V vs. Ag/AgCl. The HPLC separation of oxalate was carried out using a Capcell Pak C8 column and an isocratic mobile phase containing 80 mM KH2PO4 and 5 mM tetra-n-butylmammonium phosphate as an ion-pair reagent. The IMER was active and stable in the mobile phase employed. The plot of peak height against concentration of oxalate was linear in the range 0.1-1.6 mumol/ml with a correlation coefficient of 0.9989. The detection limit was 10 nmol/ml at a signal-to-noise ratio of 3. The within-day and between-day coefficients of variation were 5.3 and 7.9%, respectively.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.