Abstract
A microbial sensor consisting of immobilized Trichosporon brassicae, a gas-permeable Teflon membrane and an oxygen electrode is suitable for the continuous determination of acetic acid in fermentation broths. When an acetic acid solution is pumped through the flow system, the current decreases to a steady state with a response time of 8 min; shorter pumping times give peaks which can also be measured. The relationship between the current decrease and the acetic acid concentration is linear up to 54 mg l -1, with a relative standard deviation of about 6% at the higher concentrations. Selectivity is satisfactory. Results obtained with this sensor and by gas chromatography for a glutamic acid fermentation broth were in good agreement (regression coefficient 1.04). The sensor was stable for more than 3 weeks and 1500 assays.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.