Abstract

The analytical detection of d-glucose by means of self-catalytic carbon paste electrodes is described. In the construction of these electrodes, carbon powder has been modified with the redox liquid n-butylferrocene, which simultaneously serves both to help bind the paste together whilst also acting as a mediator in the enzymatic oxidation of d-glucose by glucose oxidase. The sensor then functions by monitoring the electrochemical oxidation of the constituent n-butylferrocene itself. Through testing in model glucose solutions, the electrodes were found to yield a linear response over a d-glucose concentration range of 2-20 mM. They were also successfully employed in the determination of d-glucose levels in a spiked blood sample, giving a detection limit of 0.8 mM (based on the 3sigma criterion).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.