Abstract

Single crystal MgO specimens having low load Vickers indentations were thinned in an ion milling machine employing a single ion gun, and their characteristics were investigated with optical microscopy and high voltage electron microscopy (HVEM). It was found that the state of cleanliness of the specimen chamber of the ion milling machine had a very marked influence on the quality of the thinned specimens. If the specimen chamber was not well cleaned before ion milling a fresh specimen, the latter tended to show amorphisation due to the deposition on the specimen of the debris left in the chamber from the previously ion-milled specimens. Such observations were made from MgO specimens ion milled in several different types of commercial ion milling machine employing a single gun. It is proposed that to obtain good-quality ion milled TEM specimens, it is important to clean the specimen chamber thoroughly prior to milling.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.