Abstract

Polycrystalline ammonium dithionate has been examined for its radiation response in the low dose range (<5 Gy) using EPR technique. The SO 3 − radical ion was detected as a single EPR line with a peak-to-peak derivative width of ca. 0.44 mT in irradiated samples and its intensity was found to vary linearly with dose. At equal and moderate settings of microwave power and modulation amplitude ammonium dithionate was at least seven times more sensitive than l-alanine which is the most common EPR dosimeter standard. Pulse experiments were performed on the powder samples to obtain the longitudinal relaxation time. These and microwave saturation experiments served to indicate the optimal microwave power to be applied during measurements as an EPR dosimeter for best sensitivity of this material. It is thus claimed that ammonium dithionate has excellent potential to become an EPR dosimeter with a low limit of the measurable dose for cases where tissue equivalence is not required or can be corrected for.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.