Abstract

The determination of amino acids has been achieved by reverse-phase high-pressure liquid chromatography of their dinitrophenyl derivatives. The methods developed permit the quantitation of all amino acids commonly encountered in a protein hydrolysate and the effect of various parameters on this separation was systematically evaluated. The procedure eliminates the need for specialized postcolumn equipment as employed in conventional amino acid analysis and can be obtained by a simple gradient high-pressure chromatograph. The sensitivity obtained is comparable to that available by methods in common usage, being able to determine amino acids quantitatively in the low picomole range.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.