Abstract

Our study aimed to compare the accuracy of step count and ambulation distance determined with the Yamax Digi-Walker SW-700 pedometer (DW) and the Ossur patient activity monitor (PAM) in 20 transtibial amputation subjects who were functioning at the K3 Medicare Functional Classification Level. Subjects completed four simulated household tasks in an apartment setup and a gymnasium walking course designed to simulate outdoor walking without the presence of environmental barriers or varied terrain. The mean step count accuracy of the DW and the PAM was equivalent for both the household activity (75.3% vs 70.6%) and the walking course (93.8% vs 94.0%). The mean distance measurement accuracy was better with the DW than with the PAM (household activity: 72.8% vs 0%, walking course: 92.5% vs 86.3%; p < 0.05). With acceptable step count accuracy, both devices are appropriate for assessing relatively continuous ambulation. The DW may be preferred for its more accurate distance measurements. Neither device is ideal for monitoring in-home ambulation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.