Abstract

For temperature measurements of targets at low temperatures, dual-band radiation thermometry using mid-infrared detectors has been investigated extensively. However, the accuracy is greatly affected by the reflected ambient radiation and stray radiation, which depend on the ambient temperature. To ensure measurement accuracy, an improved dual-band measurement model is established by considering the reflected ambient radiation and the stray radiation. The effect of ambient temperature fluctuation on temperature measurement is then further analyzed in detail. Experimental results of measuring a gray-body confirm that the proposed method yields high accuracy at varying ambient temperatures. This method provides a practical approach to remove the effect of ambient temperature fluctuations on temperature measurements.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.