Abstract

Readings of hermetically sealed force transducers are affected by fluctuations in atmospheric pressure, and measured deflections vary when the ambient pressure changes quickly between two readings at zero force and at a certain force step. For some types of force transducers, readings at the zero point are quite sensitive to the ambient pressure change and this influence is not negligible. By periodically repeating measurements on two force transducers, it was verified that the repeatability of the measurements can be improved, and hence calibration uncertainty can be reduced to some extent, by compensating for deflections of the force transducers due to ambient pressure fluctuations by referring to the zero sensitivity.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.