Abstract

The instrument is based on a miniature plasma source mounted at an oblique angle close to the injection gate of the ion mobility spectrometer. The plasma torch consists of two 5 mm wide external cylindrical electrodes, 10 mm apart, which are placed coaxially around a fused silica tube (1.5 mm i.d. and 3.0 mm o.d.). A small helium plasma is created by applying a alternating voltage of 8 kV at 28 kHz and employed for the direct desorption and ionization of solid or liquid samples, which are placed on an electrically isolated support. The separation section of the ion mobility spectrometer has a drift tube of 10 cm length and an applied high voltage of 4 kV. The instrument was built in-house at low cost and can easily be duplicated. Its usefulness was demonstrated by the rapid identification of five different pharmaceutical drugs, namely acetaminophen, loratadine, norfloxacin, tadalafil, thiamine as well as caffeine in ground coffee beans.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.