Abstract

Gas-flow proportional counting systems are used by the Radiation Metrology Laboratory (RML) at Sandia National Laboratories for reactor fluence monitoring with the 32S(n,p)32P reaction. Calibration of these systems has traditionally been accomplished by fluence-transfer irradiations at the NIST 252Cf facility. Such calibrations have become increasingly difficult as the NIST 252Cf source decayed to unusable levels. To minimize the risk to the testing programs from an inability to properly calibrate these systems, the RML has developed two alternative calibration techniques: 1) development and implementation of certified 32P sources for activity calibrations and subsequent calculation of neutron fluence, and 2) direct counting of non-certified reactor-irradiated sulfur pellets by liquid scintillation counting to determine 32P activity for the subsequent calibration of gas-flow proportional counters. Preliminary comparisons show that the several calibration methods are capable of overall uncertainties within about 5 percent.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.