Abstract
To compare the diagnostic efficacy of wide-field digital retinal imaging (WFDRI) with binocular indirect ophthalmoscopy (BIO) for retinopathy of prematurity (ROP) screening. Premature infants admitted for ROP screening were included in this prospective, randomized and double-blind comparative study. They were examined by using BIO and WFDRI. A total of 58 infants were enrolled in the study. The sensitivities of WFDRI in detecting any stage of ROP, treatment-requiring ROP and plus disease were 58.6, 100, and 100% respectively, with a specificity of 100% for all. The proportional agreement between WFDRI and BIO was 0.903 for detection of any stage of ROP, 1.0 for treatment-requiring ROP, and 1.0 for plus disease. The sensitivity and specificity of WFDRI was excellent for the diagnosis of severe and treatment-requiring ROP. However, BIO was superior in mild ROP particularly for the ones in retinal periphery.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.