Abstract

An alternative and practical method is described for simultaneous detection and quantification of the potent hallucinogen lysergic acid diethylamide (LSD) and related compounds in urine and serum samples. The procedure is based on liquid–liquid re-extraction with ethyl acetate and reversed-phase liquid chromatography coupled with fluorescence detection (HPLC–FLD). With detection limits in urine and serum samples of ca 0.07 ng mL−1 for LSD, nor-LSD, and iso-LSD, respectively, the method is well suited to forensic investigations. Application of the method to clinical samples and autopsy material enable selective identification and accurate quantification of LSD and related compounds. Comparison of results with those obtained from an LSD immunoassay (EMIT II) emphasize the need for chromatographic confirmation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.