Abstract

This paper addresses alternative procedures to the ESO supplied pipeline procedures for the reduction of UVES spectra of two quasar spectra to determine the value of the fundamental constant μ=Mp/Me at early times in the universe. The procedures utilize intermediate product images and spectra produced by the pipeline with alternative wavelength calibration and spectrum addition methods. Spectroscopic studies that require extreme wavelength precision need customized wavelength calibration procedures beyond that usually supplied by the standard data reduction pipelines. An example of such studies is the measurement of the values of the fundamental constants at early times in the universe. This article describes a wavelength calibration procedure for the UV–visual Echelle spectrometer on the very large telescope, however, it can be extended to other spectrometers as well. The procedure described here provides relative wavelength precision of better than 3×10-7 for the long-slit Thorium–Argon calibration lamp exposures. The gain in precision over the pipeline wavelength calibration is almost entirely due to a more exclusive selection of Th/Ar calibration lines.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.