Abstract
A novel fiber-optic polarimetric configuration for a magnetometer utilizing the Faraday effect is described. The system is operated in a closed-loop mode using magneto-optic feedback through a simple current-driven solenoid to maintain the azimuth of the optical output from the polarimeter constant. In this mode of operation scale-factor errors due to changes in the Verdet constant of the fiber are effectively eliminated. The minimum detectable current was less than 1 mA, with linear operation up to several amperes. A multiplexed system was also demonstrated, with potential application as a multiaccess noninvasive data ring.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.