Abstract

An atmospheric pressure plasma sustained by an alternating-current power supply was generated using the electrolyte solution as one electrode. The emission spectra of the plasma are related to the metal ions present in the solution and thus can be used for metal ion detection. The main advantage of this source is that it operates at low flow rates in the range of 0.1 to 0.8 mL/min. The effects of the discharge electrolyte anion, the pH and solution flow rate on emission signal were studied. The analytical response curves for Na and Cd demonstrated good linearity at least in the range of 0.5–100 mg L−1. The limit of detection of Na and Cd were determined to be 0.04 and 0.09 mg L−1, respectively. The method was applied for the determination of elements in simulated natural water samples (GBW(E)080408 and GBW(E)080402). The obtained results were in good agreement with the reference values. It provides an alternative analytical method for the determination of elements in water samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.