Abstract

The accuracy with which the concentration dependence of the sedimentation coefficient, s=s 0(1 - kc), can be quantified for globular proteins by commonly used procedures has been examined by subjecting simulated sedimentation velocity distributions for ovalbumin to c(s)‒s analysis. Because this procedure, as well as its g(s)‒s counterpart, is based on assumed constancy of s over the time course of sedimentation coefficient measurement in a given experiment, the best definition of the concentration coefficient k is obtained by associating the measured s with the mean of plateau concentrations for the initial and final distributions used for its determination. The return of a slightly underestimated k (by about 3%) is traced to minor mislocation of the air‒liquid meniscus position as the result of assuming time independence of s in a given experiment. Although more accurate quantification should result from later SEDFIT and SEDANAL programs incorporating the simultaneous evaluation of s 0 and k, the procedures based on assumed constancy of s suffice for determining the limiting sedimentation coefficient s 0-the objective of most s‒c dependence studies.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.