Abstract
This work investigates a disposable creatinine potentiometric biosensor based on a pH-sensitive electrode and an ammonium ion-selective electrode. An enzyme, creatinine iminohydrolase (CIH) was immobilized directly onto the surface of the electrodes by the entrapment method. According to the experimental results, the creatinine biosensors based on the pH-sensitive electrode and ammonium ion-selective electrode show a rapid response time (t95 < 30 s) and a linear dynamic response range for creatinine determination between 40 muM and 140 muM. Furthermore, the ammonium ion-selective electrode-based biosensors exhibited good operational stability (over 14 times) and storage stability (over 40 days). By discussing the reproducibility and pH effects of the creatinine biosensors, features of creatinine biosensors based on ammonium ion-selective electrodes are described. Moreover, development of a low-cost, solid-state, disposable potentiometric creatinine biosensor based on pH-sensitive electrodes and ammonium ion-selective electrodes is discussed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.