Abstract

A variant of the scheme of a magnetometric sensor based on cesium atomic vapor is proposed and experimentally investigated. The sensor uses magnetic resonance excitation by modulated light of an hyperfine optical pumping that is transverse to the magnetic field. It is shown that, for a cell with a volume of 0.125 cm3, the variational sensitivity of such a scheme, estimated from the ratio of the steepness of the signal in the center of the magnetic resonance to the shot noise of the detecting radiation, reaches a level of <10 fT/Hz1/2 in the frequency band width of the order of 850 Hz. The sensor, which does not emit radio frequency fields, is designed for use in magnetoencephalographic complexes. Possible ways to improve the performance of the scheme for detecting relatively fast (~4.2 kHz in a field of 0.1 mT) signals of the precession of proton magnetic moments in promising ultra-weak field tomography schemes are considered.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.