Abstract

We describe a purification system based on a commercial preparative-scale gas chromatograph with a custom-designed condenser, collector, and fraction handling system. In our fraction collector design, all the wetted surfaces were either 316 stainless-steel or nickel. The collectors and the integrated gas-handling manifold were designed to be used down to liquid nitrogen temperature and up to 7 MPa of pressure to accommodate low-boiling-point compounds, such as refrigerants. The design, operation, and performance of this apparatus are presented.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.