Abstract

A new Al(III) selective membrane based on 5-bromo(salicylidene-2-amino thiophenol) as a neutral carrier in a poly(vinyl chloride) (PVC) matrix is coated on platinum wire to determine trace amounts of Al(III) ion in various samples. The sensor exhibits a good response for Al(III) ion over a wide concentration range of 5.20 × 10-7 to 1.00 × 10-2 M with Nernstian slope of 19.64 ± 0.15 mV decade-1 and low detection limit of 5.01 × 10-7 M. The performance of the sensor is best in the pH range of 4.00-6.50 without any divergence in potential and it also works well in partially non-aqueous medium. The coated wire electrode has a short response time of about 5 s and is stable for at least 2 months. The proposed membrane electrode revealed good selectivity for Al(III) ion over a wide variety of other metal ions. It was successfully applied for determination of Al(III) ion in different real and environmental samples and also as an indicator electrode for potentiometric titration of Al(III) ion with EDTA.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.