Abstract

Signal averaged atrial activity recorded from the body surface is finding increased use in assessing atrial arrhythmias. The process of signal averaging is critically dependent upon precise alignment of the waveform of interest. The alignment accuracy of wide (0.01-300 Hz) and narrow (10-300 Hz) bandpass filtered P waves recorded from the body surface were compared to atrial electrograms recorded from temporary pacing wires. The locations of the filtered P waves relative to the atrial electrogram were calculated beat-by-beat for 100 beats in 15 patients aged 1 month to 71 years. The standard deviation of the alignment error for the wide filtered surface P wave averaged 1.7 +/- 0.8 msec and for the narrow filtered P waves was 0.8 +/- 0.4 ms. These results demonstrate that high pass filtering will improve alignment accuracy, and therefore, improve frequency resolution for the analysis of signal averaged P waves.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.