Abstract
Recently a high-performance PET scanner, which measures depth-of-interaction (DOI) information, is being developed for molecular imaging. DOI measurement of multi-layered thin crystals can improve spatial resolution and scanner sensitivity simultaneously. In this paper, we apply an algebraic image reconstruction method to 2-dimensional (2D) DOI-PET scanners using accurate system modeling, in order to evaluate the effects of using DOI information on PET image quality. Algebraic image reconstruction methods have been successfully used to improve PET image quality, compared with the conventional filtered backprojection method. The proposed method is applied to simulated data for a small 2D DOI-PET scanner. The results show that accurate system modeling improves spatial resolution without noise emphasis, and that DOI information improves uniformity of spatial resolution.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Japanese Journal of Medical Physics (Igakubutsuri)
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.