Abstract

Abstract. A four-channel thermal dissociation laser induced fluorescence (TD-LIF) instrument has been developed for simultaneous measurements of nitrogen dioxide (NO2), total peroxy nitrate (∑PNs), total alkyl nitrate (∑ANs) and nitric acid (HNO3). NO2 is measured directly by LIF at 532 nm, whereas organic nitrates and nitric acid are thermally dissociated at distinct temperatures in the inlet to form NO2, which is then measured by LIF. The concentrations of each dissociated species are derived by the differences in measured NO2 relative to the reference colder inlet channel. The TD-LIF was adapted to fly on board the UK Facility for Airborne Atmospheric Measurements (FAAM) BAe 146-301 atmospheric research aircraft in summer 2010, and to date has successfully flown in five field campaigns. This paper reports novel improvements in the TD-LIF instrumentations, including (1) the use of a single wavelength laser, which makes the system compact and relatively cheap; (2) the use of a single beam laser that allows easy alignment and optical stability against the vibrational aircraft environment; and (3) the optical assembly of four detection cells that allow simultaneous and fast (time resolution up to 0.1 s) measurements of NO2, ∑PNs, ∑ANs and HNO3. Laboratory-generated mixtures of PNs, ANs and HNO3 in zero air are converted into NO2 and used to fix the dissociation temperatures of each heated inlet to test the selectivity of the instrument and potential interferences due to recombination reactions of the dissociated products. The effectiveness of the TD-LIF was demonstrated during the RONOCO aircraft campaign (summer 2010). A chemiluminescence system that was measuring NO2 and a broadband cavity enhanced absorption spectrometer (BBCEAS) that was measuring one of the PNs (N2O5) were installed on the same aircraft during the campaign. The in-flight intercomparison of the new TD-LIF with the chemiluminescence system for NO2 measurements and the intercomparison between ∑PNs measured by the TD-LIF and N2O5 by the BBCEAS are used to assess the performance of the TD-LIF.

Highlights

  • System compact and relatively cheap; (2) the use of a single beam laser that allows easy alignment and optical stability against the vibrational aircraft environment; and (3) the op

  • During this test a synthetic mixture of about 0.9 ppbv of peroxyacetyl nitrate (PAN) and 2.4 ppbv of ethyl nitrate was supplied to the common inlet of the thermal dissociation laser induced fluorescence (TD-laser induced fluorescence (LIF))

  • The results of NO3Comp 2007 intercomparison exercise, carried out in the SAPHIR chamber, proved that three cavity ring-down instruments and two LIF instruments detect N2O5 with good agreement (Fuchs, et al, 2012); here we have reported the intercomparison in troposphere on board the BAE 146-301 aircraft

Read more

Summary

The TD-LIF Instrument

The TD-LIF uses the laser induced fluorescence technique with a single-wavelength laser for direct detection of NO2, which has previously been applied in ground-based systems (Matsumoto et al, 2001; Dari-Salisburgo et al, 2009).

Optical layout and LIF system
Sampling and inlet system
Data retrieval and calibration
Laboratory tests: zeroing and selectivity
Comparison of in-flight measurements of TD-LIF
Laboratory tests
Peroxy nitrates
Conclusions