Abstract

In the medical and nuclear fields, there are environments where exposure to photons with energies above several MeV can result in problems. The National Metrology Institute of Japan has developed a high-energy photon field using a 15 MeV electron beam of a clinical linear accelerator with a copper target and an aluminium filter unit to facilitate dosimeter calibration in terms of air kerma. To determine the air kerma rate, the energy fluence distribution at a reference point was calculated, and both calculations and experiments evaluated the effective energy and spatial dose distribution. Moreover, to validate the air kerma measurement, two commercial cavity chambers were calibrated in a developed photon field. The results obtained exhibited a 4% difference compared with those in a Co-60γ-ray reference field.