Abstract

PurposeFew studies have evaluated real-world performance of radiological AI-tools in clinical practice. Over one-year, we prospectively evaluated the use of AI software to support the detection of intracranial large vessel occlusions (LVO) on CT angiography (CTA). MethodQuantitative measures (user log-in attempts, AI standalone performance) and qualitative data (user surveys) were reviewed by a key-user group at three timepoints. A total of 491 CTA studies of 460 patients were included for analysis. ResultsThe overall accuracy of the AI-tool for LVO detection and localization was 87.6%, sensitivity 69.1% and specificity 91.2%. Out of 81 LVOs, 31 of 34 (91%) M1 occlusions were detected correctly, 19 of 38 (50%) M2 occlusions, and 6 of 9 (67%) ICA occlusions. The product was considered user-friendly. The diagnostic confidence of the users for LVO detection remained the same over the year. The last measured net promotor score was −56%. The use of the AI-tool fluctuated over the year with a declining trend. ConclusionsOur pragmatic approach of evaluating the AI-tool used in clinical practice, helped us to monitor the usage, to estimate the perceived added value by the users of the AI-tool, and to make an informed decision about the continuation of the use of the AI-tool.