Abstract

ObjectiveTo assess the agreement of sleep parameters measured by two actigraphs (SOMNOwatch plus®, ActiGraph GT3X+®) at two different placements (wrist, hip) and of self-reported sleep with polysomnography (PSG). MethodsWe estimated agreement with PSG for total sleep time (TST), sleep onset latency (SOL), wake after sleep onset (WASO), number of awakenings after sleep onset (NASO), and sleep efficiency (SE%) for 100 participants of the general population, aged 18–75 years by judging mean differences to PSG and intervals of agreement using Bland–Altman plots. ResultsMean difference to PSG for TST was 8.3 min (95% confidence intervals [CI] −7.4; 24.1) for SOMNOwatch plus® (wrist), 39.8 min (95% CI 24.3; 55.3) for self-report, −79.0 min (95% CI −89.0; −68.9) for SOMNOwatch plus® (hip), and −81.1 min (95% CI −91.9; −70.4) for GT3X+® (hip), respectively. The width of intervals of agreement differed with the placement of the devices. Mean differences to PSG were higher for hip-based measurements compared with wrist placement for most parameters. ConclusionsAgreement of sleep parameters assessed by actigraphy with PSG differs with the placement of the device and is limited for hip-based measurements. Agreement of self-report with PSG is comparable to that of actigraphy for some parameters.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call